Boonton 4500b rf peak power analyzer, Operation – Boonton 4500B Peak Power Meter User Manual

Page 96

Advertising
background image

Boonton 4500B RF Peak Power Analyzer

Operation

4-24

Table 4-4B. Chan # > Extensions > Define Pulse > Submenu (continued)

Menu Item

Function

(Type)

Selections

(SCPI cmd)

HINT

If you want to use 10% and 90% reference levels in measuring risetime, and you want to use the rule of
thumb relationship between risetime and bandwidth, always set Pulse Units to Volts. If you also want to
view a waveform that most closely resembles an oscilloscope display of the same signal, set the display
units to Volts. The display units setting is optional and will not affect the automeasure result.

Start Gate

Pulse

Sets the beginning of the pulse measurement region as a percentage of

(Numeric)

Continuous range:

pulse width

0.00 % to 40.00 %

SENSe[1|2|3|4|6|7]:PULSe:STARTGT <0.00 to 40.00> (percent)

.
End Gate

Pulse

Sets the end of the pulse measurement region as a percentage of pulse

(Numeric)

Continuous range:

width

60.00 % to 100.00 %

SENSe[1|2|3|4|6|7]:PULSe:ENDGT <60.00 to 100.00> (percent)


























Fig. 4-17b. Markers Indicate Channel 1 power.

Fig. 4-17c. Markers Indicate Channel 2 power.

Advertising