Boonton 4500B RF Peak Power Analyzer

Operation

4-75

Table 4-22. Util > Remote Conf > Emulation > Submenu


Menu Item

Function

(Type)

Selections

(SCPI CMD)



Boot Language

SCPI, 4500A

Boot Language selects the default language

(Toggle)

at instrument power-up. It does not change

the Current Language.



Curr Language

SCPI, 4500A

Current Language permits viewing and

(Toggle)

changing the language at any time without a

remote command. At power-up, the Current

Language is set to the Boot Language.


SYSTem:LANGuage EMUL4500A



*IDN Response

4400, 4400A

Selects model number the instrument returns

(Mult. Choice)

4500, 4500A

in Emulation mode to *IDN? query.

4500B



Option Select

None, -S/1,

Selects the option identifier appended to

(Mult. Choice)

-04, -05, 06

the Emulation mode *IDN? query.



Ignore Unsupp

On, Off

In 4400/4500/4400A/4500AEmulation mode

(Toggle)

only, if ―On‖, ignores GPIB bus when an
unsupported command is received.
If ―Off‖, returns error message.





Fig. 4-38. Util >

Remote Conf >

Emulation > submenu

Note

For compatibility purposes, the 4500B remote interface can be set to emulate the BOONTON 4400, 4500, 4400A

(including 4400A-L) and 4500A (including 4500A-L) series of instruments. For additional discussion of this

subject, see Section 5.2. To use the 4500B SCPI interface, set the Util > Remote Conf > Emulation > Curr

Language menu selection to SCPI. To emulate the 4400/4500 legacy interface, set the Util > Remote Conf >

Emulation > Curr Language menu selection to the desired model. To control the language in use at instrument

power-up, select your choice in the in the Util > Remote Conf > Emulation > Boot Language menu box.