Trigger:btrigger:slope, Trigger:btrigger:source, Trigger:btrigger:tdly – Boonton 4500B Peak Power Meter User Manual

Page 291: Trigger:btrigger:slope -91, Trigger:btrigger:source -91, Trigger:btrigger:tdly -91, Boonton 4500b rf peak power analyzer, Remote operation

Advertising
background image

Boonton 4500B RF Peak Power Analyzer

Remote Operation

5-91

TRIGger:BTRIGger:SLOPe

Description:

Set or return the B trigger slope or polarity. When set to POS, B triggers will be generated when a
signal‘s rising edge crosses the B trigger level threshold. When NEG, B triggers are generated on
the falling edge of the pulse.

Syntax:

TRIGger:BTRIGger:SLOPe <asc>

Argument:

<asc> = {NEG, POS}

Valid Modes:

Pulse and gated Statistical modes only

TRIGger:BTRIGger:SOURce

Description:

Set or return the trigger source used for the B trigger. The CH1 and CH2 (dual channel models
only) settings use the signal from the associated sensor. TRIG1 and TRIG2 (dual channel models
only) use the signal applied to the associated external trigger input.

Syntax:

TRIGger:BTRIGger:SOURce <asc>

Argument:

<asc> = {CH1, CH2, TRIG1, TRIG2}

Valid Modes:

Pulse and gated Statistical modes only

TRIGger:BTRIGger:TDLY

Description:

Set or return the B trigger arming delay time in seconds with respect to the trigger. Active only
when TRIGger:BTRIGger:MODE is set to TDLY.

Syntax:

TRIGger:BTRIGger:DELay <n>

Argument:

<n> = 0.0 – 1.0 seconds in 10 nsec increments

Valid Modes:

Pulse and gated Statistical modes only

Advertising