Display:trace:pdfbars, Display:trace:type, Display:trace:vcenter – Boonton 4500B Peak Power Meter User Manual

Page 231: Display:trace:vscale, Display:trace:pdfbars -31, Display:trace:type -31, Display:trace:vcenter -31, Display:trace:vscale -31, Boonton 4500b rf peak power analyzer, Remote operation

Advertising
background image

Boonton 4500B RF Peak Power Analyzer

Remote Operation

5-31

DISPlay:TRACe:PDFBars

Description:

Set or return the number of bars in the PDF display.

Syntax:

DISPlay:TRACe[1|2]:PDFBars <asc>

Argument:

<asc> = {1, 2, 5, 10, 20, 35}

Valid Modes:

Statistical PDF mode only.

DISPlay:TRACe:TYPE

Description:

Set or return the trace drawing style, points, lines or off. For Points, display pixels are plotted
individually. For Lines, display pixels are connected with vertical lines. For Off pixels are not
plotted on the display, however, measurements continue to be made.

Syntax:

DISPlay:TRACe[1|2|3|4|5|6|7]:TYPE <asc>

Argument:

<asc> = {POINTS, LINES, OFF}

Valid Modes:

All

DISPlay:TRACe:VCENTer

Description:

Set or return the power or voltage level of the horizontal centerline of the graph for the specified
channel in channel units. If a change in the vertical scale causes the center maximum value to be
exceeded, the center will be forced to the maximum value for the new range.

Syntax:

DISPlay:TRACe[1|2|3|4|5|6|7]:VCENTer <n>

Argument:

<n> = -200.00 to +200.00 dBm for dBm units

10,000 times the vertical scale power/div for watts units

10,000 times the vertical scale volts/div for volts units

Valid Modes:

All

DISPlay:TRACe:VSCALe

Description:

Set or return the power or voltage vertical sensitivity of the trace display in channel units.

Syntax:

DISPlay:TRACe[1|2|3|4|5|6|7]:VSCALe <n>

Argument:

Units = dBm, dBV, dBmV, dBuV , <n> = range in dB/division

Units = watts

<n> = range in watts/division

Units = volts

<n> = range in volts/division

Valid Modes:

All