Display:decades, Display:envelope, Display:fullscale – Boonton 4500B Peak Power Meter User Manual
Page 219: Display:grid, Display:decades -19, Display:envelope -19, Display:fullscale -19, Display:grid -19, Boonton 4500b rf peak power analyzer, Remote operation
Boonton 4500B RF Peak Power Analyzer
Remote Operation
5-19
DISPlay:DECAdes
Description:
Set or return the number of decades displayed on a Statistical mode (CDF, CCDF, PDF)
logarithmic horizontal scale.
Syntax:
DISPlay:DECAdes <n>
Argument:
<n> = 1 to 7
Valid Modes:
Statistical modes only
DISPlay:ENVElope
Description:
Set or return the status of the trace envelope display mode. Envelope mode plots the maximum
and minimum value for each measurement pixel.
Syntax:
DISPlay:ENVElope <asc>
Argument:
<asc> = {ON, OFF}
Valid Modes:
Any
DISPlay:FULLscale
Description:
Set or return the fullscale (right edge) value of the Statistical mode (CDF, CCDF, PDF)
logarithmic horizontal scale in percent.
Syntax:
DISPlay:FULLscale <n>
Argument:
<n> = {100, 10, 1, 0.1, 0.01}
Valid Modes:
Statistical modes only
DISPlay:GRID
Description:
Set or return the type of axes and graticule displayed in the traces box.
Syntax:
DISPlay:GRID <asc>
Argument:
<asc> = {BOX (box only), XHAIR (box with crosshair), GRID (box with grid), HASH (box with
hash-marks), XHAIRGRID (box with grid and crosshair)}
Valid Modes:
All