Table 4-32. calibrator > pulse (preset) > submenu, Boonton 4500b rf peak power analyzer, Operation – Boonton 4500B Peak Power Meter User Manual

Page 156

Advertising
background image

Boonton 4500B RF Peak Power Analyzer

Operation

4-84

Table 4-31. Calibrator > Menu (continued)

Menu Item

Function

(Type)

Selections

(SCPI cmd)

to set the Set Level value greater than the
Calibrator > Max Power value.

OUTPut:LEVel:LIMit <-50.0 to 20.0>

Cal Mode

CW, Pulse

Selects the calibrator output signal format.

(Toggle)

Press the Calibrator > Cal Mode menu key to alternately select a CW
or a Pulse modulated RF output signal.

OUTPut:MODUlation {CW, PULSE}


Pulse

MENU

Opens the Calibrator > Pulse submenu

(Submenu)


Extensions

MENU

Opens the Calibrator > Extensions submenu

(Submenu)



Table 4-32. Calibrator > Pulse (Preset) > Submenu

Menu Item

Function

(Type)

Selections

(SCPI cmd)


Source

Int, Ext

Selects the source for the calibrator output pulse modulation

(Toggle)

Press the Calibrator > Pulse > menu key to specify whether the
calibrator output pulse is to be internally or externally generated.


OUTPut:PULSe:SOURce {INT, EXT}



Polarity

+,

Selects the polarity for calibrator’s preset internal pulse modulation

(Toggle)

Press the Calibrator > Pulse > Polarity menu key to select positive-
going or negative-going calibrator output pulses when Preset internal
pulse modulation is used. This control has no effect on External or
Internal Variable pulse modulation.

OUTPut:POLArity {POS, NEG}