Table 4-11. mark > menu, Table 4-11. mark > menu -52, Boonton 4500b rf peak power analyzer – Boonton 4500B Peak Power Meter User Manual

Page 124: Operation

Advertising
background image

Boonton 4500B RF Peak Power Analyzer

Operation

4-52

Procedure

To make time interval measurements, position Time Marks 1 and 2 at the beginning and end of the interval

you wish to measure. The time interval measurement appears in the Mark > Delta Time window. The

resolution of time interval measurements depends on the timebase selection (see Subsection 4.10).



Table 4-11. Mark > Menu

Menu Item

Function

(Type)

Selections

(SCPI cmd)


Marker #

Pulse & Mod

Controls the position of Marker # in time units.

(Numeric)

Continuous Range:

(Left and Right

Press the Mark > Marker # menu key to

boundaries of the activate this function.

display.)

Use any of the data entry controls to position
the time marker anywhere within the left and
right boundaries of the waveform display. The
marker position is expressed as the time offset
relative to the trigger event.


Stat

Controls the position of Marker # in percent.

Continuous Range:

(Left and Right

Press the Mark > Marker # menu key to

boundaries of the

activate this function.

display.)

Use any of the data entry controls to position
the percent marker anywhere within the left
and right boundaries of the distribution
display. The marker position is expressed as
the percent of total occurrence.


MARKer[1/2]:POSItion:TIMe <arg> (sec)


MARKer[1/2]:POSItion:PERcent

<0 thru 100> (percent)


MARKer[1/2]:POSItion:PIXel <0 thru 500>



Delta Time

Pulse & Mod

Displays the time difference between

(None)

Marker 1 and Marker 2


The value that appears in this window is

automatically calculated by the instrument. Fig. 4-28. Mark > menu

Stat

Displays the percent difference between
Marker 1 and Marker 2

The value that appears in this window is
automatically calculated by the Model 4500B.

Advertising