Display:pctofst, Display:pctscale, Display:persistence – Boonton 4500B Peak Power Meter User Manual

Page 227: Display:plabel:mode, Display:pctofst -27, Display:pctscale -27, Display:persistence -27, Display:plabel:mode -27, Boonton 4500b rf peak power analyzer, Remote operation

Advertising
background image

Boonton 4500B RF Peak Power Analyzer

Remote Operation

5-27

DISPlay:PCTOFST

Description:

Set or return the horizontal trace percent offset of the statistical linear display mode.

Syntax:

DISPlay:PCTOFST <n>

Argument:

<n> = 0.0 to 99.0 % with 0.1% resolution (see restrictions)

Valid Modes:

Statistical mode only

Restrictions:

The sum of (PCTSCALE * 10 div) and PCTOFST must not be greater than 100%.

DISPlay:PCTSCALE

Description:

Set or return the horizontal trace percent/div scale in the statistical linear display mode. Range
settings are fixed in a 1, 2, 5 sequence, and if the argument does not match one of the settings, it
will be forced to the next highest entry.

Syntax:

DISPlay:PCTSCALE <n>

Argument:

<n> = {0.1, 0.2, 0.5, 1, 2, 5, 10} %

Valid Modes:

Statistical mode only

Restrictions:

The sum of 10*PCTSCALE and PCTOFST must not be greater than 100%.

DISPlay:PERSistence

Description:

Set or return the status of the display persistence function. Affects CH1, CH2, TRIG1, TRIG2 and
CHM. The ON argument sets the persistence mode and forces the dot trace type. See
DISPlay:TRACe:TYPE.

Syntax:

DISPlay:PERSistence <asc>

Argument:

<n > = {0, 1}

Valid Modes:

All

DISPlay:PLABel:MODE

Description:

Set or return the status of the plot label display function. The plot label consists of 4 user
programmable text lines in the trace box.

Syntax:

DISPlay:PLABel:MODE <asc>

Argument:

<asc > = {ON, OFF}

Valid Modes:

All