Trg -10, Tst? -10, Wai -10 – Boonton 4500B Peak Power Meter User Manual

Page 210: Boonton 4500b rf peak power analyzer, Remote operation

Advertising
background image

Boonton 4500B RF Peak Power Analyzer

Remote Operation

5-10

*TRG

Description:

Simulate the bus trigger command.. This command has the same effect as GET, Group Execute
Trigger, except it must be parsed and decoded before the action takes place. There is no query
form of this command.

Syntax:

*TRG

Argument:

None

Valid Modes:

Any


*TST?

Description:

Self-test query. This command initiates a self-test of the instrument, and returns a result code
when complete. The result is zero for no errors, or a signed, 16-bit number if any errors are
detected.

Syntax:

*TST

Returns:

Error Code

Valid Modes:

Any

*WAI

Description:

Wait command. This command insures sequential, non-overlapped execution. The 4500B always
operates in non-overlapped, sequential mode. The command is accepted as valid but takes no
action.

Syntax:

*WAI

Argument:

None

Valid Modes:

Any

Advertising