Boonton 4500B RF Peak Power Analyzer

Remote Operation

5-65

REFLine:TRACk

Description:

Controls the automatic tracking of waveform parameters using reference lines. The reference lines
must be assigned to the measurement channel of interest. For CHM, MEM1 and MEM2, only the
marker tracking function X1X2 is available.

Syntax:

REFLine:TRACk <asc>

Argument

<asc> = {OFF, X1X2, TOPBOT, DISTMESIAL, DISTPROX}

OFF

No tracking. Reference lines have static positions

X1X2

Reference lines track the intersections, respectively, of Marker 1 and Marker 2
with the assigned channel‘s waveform.

TOPBOT

Reference lines track the Top and Bottom level of the assigned channel‘s
waveform.

DISTMESIAL

Reference lines track the Distal and Mesial level of the assigned channel‘s
waveform.

DISTPROX

Reference lines track the Distal and Proximal level of the assigned channel‘s
waveform

The terms Top Level, Bottom Level, Distal, Proximal and Mesial level are those set in the Pulse
Definitions menu (or by GPIB command) and refer to the IEEE Definitions of Pulse parameters.

Valid Modes:

Pulse mode.

Advertising