Boonton 4500B RF Peak Power Analyzer

Remote Operation

5-104

Table 5-1. SCPI COMMAND SUMMARY (continued)


OUTPut:PULSe:CTRL

Set/return pulse control. <arg> = PRESET, VAR


OUTPut:PULSe:DCYCle Set/return duty cycle. <arg> = 10, 20, 30, 40, 50

OUTPut:PULSe:PERWID Set/return variable calibrator pulse period and pulse width.

<arg1> = 28e-6 to 131.07e-3 <arg2> = 7e-6 to 65.535e-3


OUTPut:PULSe:PERiod

Set/return preset calibrator pulse period in millseconds. <arg> = 0.1, 1, 10


OUTPut:PULSe:SOURce

Set/return calibrator pulse modulation source. <arg> = INT, EXT


OUTPut:SIGnal

Set/return state of calibrator RF output. <arg> = ON, OFF


OUTPut:TRIGout[1/2]:SOURce

Set/return signal source for trigger output. <asc> = CH1/2, TRIG1/2, SWP, CLK


OUTPut:TRIGout[1/2]:POLarity

Set/return trigger output polarity. <asc> = POS, NEG


READ[1|2|3|4|6|7]:ARRay:AMEAsure:POWer?

Return new Ppeak, Pavgcyc, Pavgpulse, Ptop, Pbot, Overshoot


READ[1|2|3|4|6|7]:ARRay:AMEAsure:STATistical? Return new Pavg, Ppeak, Pmin, Pk/Avg, Pmrk1, Pmrk2,

PCTmrk1, PCTmrk2, Tot(Msamples)


READ[1|2|3|4|6|7]:ARRay:AMEAsure:TIMe?

Return new Freq, Period, Width, Offtime, Dcyc, Risetime, Falltime,
EdgeDly, Skew


READ[1|2|3|4|6|7]:ARRay:CW:POWer?

Return new Pavg, Pmax, Pmin, Ppulse or PK/AVG


READ[1|2|3|4|6|7]:ARRay:MARKer:POWer?

Return current Pavg, Pmax, Pmin, P/Av, Pmrk1, Pmrk2, Pmrk1/Pmrk2


READ:ARRay:MARKer:WINDow?

Return current reading at MK1, MK2, and center window.


READ[1|2|3|4|5|6|7]:CW:POWer?

Return new average reading in power units


READ[1|2|3|4|5|6|7]:INTERval:AVERage?

Return average power between MK1 and MK2.


READ[1|2|3|4|6|7]:INTERval:MAXimum?

Return maximum instantaneous power between MK1 and MK2.


READ[1|2|3|4|6|7]:INTERval:MINimum?

Return minimum instantaneous power between MK1 and MK2.


READ[1|2|3|4|6|7]:INTERval:MAXFilt?

Return maximum filtered power between MK1 and MK2.


READ[1|2|3|4|6|7]:INTERval:MINFilt?

Return minimum filtered power between MK1 and MK2.


READ[1|2|6|7]:INTERval:PKAVG?

Return peak to average power between MK1 and MK2.


READ[1|2|3|4|5|6|7]:MARKer[1|2]:AVERage?

Return current reading at the specified marker.


READ[1|2|3|4|5|6|7]:MARKer:DELTa?

Return difference between MK2 and MK1.


READ[1|2|3|4|6|7]:MARKer[1|2]:MAXimum?

Return maximum reading at the specified marker


READ[1|2|3|4|6|7]:MARKer[1|2]:MINimum?

Return minimum reading at the specified marker