Boonton 4500b rf peak power analyzer – Boonton 4500B Peak Power Meter User Manual

Page 352

Advertising
background image

Boonton 4500B RF Peak Power Analyzer

Maintenance

7-20


Table 7-9. 56518/526/540, 57518/540
Sensor Linearity (CW)

Measured

Cal Level Minimum CH1 CH2 Maximum
(dBm) (dB) (dB) (dB) (dB)


20.0 19.83 ___________ ___________ 20.17
15.0 14.83 ___________ ___________ 15.17
10.0 9.83 ___________ ___________ 10.17
5.0 4.83 ___________ ___________ 5.17
0.0 -0.17 ___________ ___________ 0.17
-5.0 -5.17 ___________ ___________ -4.83
-10.0 -10.17 ___________ ___________ -9.83
-15.0 -15.17 ___________ ___________ -14.83
-20.0 -20.17 ___________ ___________ -19.83
-25.0 -25.18 ___________ ___________ -24.82
-30.0 -30.19 ___________ ___________ -29.81
-35.0 -35.24 ___________ ___________ -34.76
-40.0 -40.39 ___________ ___________ -39.62
-45.0 -45.92 ___________ ___________ -44.19
-50.0 -53.18 ___________ ___________ -48.07


Advertising