Boonton 4500b rf peak power analyzer, Operation – Boonton 4500B Peak Power Meter User Manual

Page 108

Advertising
background image

Boonton 4500B RF Peak Power Analyzer

Operation

4-36

Table 4-7D. Ext Trig # > Extensions > Submenu (continued)

Menu Item

Function

(Type)

Selections

(SCPI cmd)


Define Pulse

Pulse & Mod

Selects pulse level criteria for pulse time measurements

(Submenu)

MENU


Identical to Chan # > Exten > Define Pulse menu. See Table 4-4.



Video BW

Pulse, Mod & Stat

Selects either the narrowband or wideband external trigger response,

(Toggle)

Low, High

Press Ext Trig # > Extensions > Video BW > Low to select a low
bandwidth trigger response that is useful for triggering on or displaying
noisy, low frequency signals..

Press Ext Trig # > Extensions > Video BW > High to select a wideband
response that is useful for triggering on or displaying high frequency
signals, clean signals of any bandwidth or short pulses.

SENSe[3|4]:BANDwidth {LOW, HIGH}


Averaging

Pulse

Selects the number of samples that are averaged at each point

(Numeric)

1 to 16,384

Press Ext Trig # > Extensions > Averaging to activate this function.
Use any of the data entry controls to select the number of samples to be
averaged at each point of the waveform to produce the waveform
display. Default = 8.

SENSe[3|4]:AVERage <1 to 16384>