Table 4-12. mark > extensions > submenu, Table 4-12. mark > extensions > submenu -54, Boonton 4500b rf peak power analyzer – Boonton 4500B Peak Power Meter User Manual

Page 126: Operation

Advertising
background image

Boonton 4500B RF Peak Power Analyzer

Operation

4-54

Table 4-12. Mark > Extensions > Submenu

Menu Item

Function

(Type)

Selections

(SCPI cmd)


MK Group

Pulse, Mod & Stat

Selects the marker channel assignment mode.

(Toggle)

Both, Each

When ―Both‖ has been selected, both markers
are assigned to the same channel. Changing
either marker channel assignment changes
both assignments. Also, selecting ―Both‖
activates two choices in the Mark > Exten >
Delta Marker
menu, ―Ratio‖ and ―Average.‖

When ―Each‖ has been selected, the markers
can be assigned individually; that is, Mark >
Extensions > Mk 1 CH
and Mark >
Extensions > Mk 2 CH
can be assigned to
different channels. The marker channel
assignments for the ―Both‖ and ―Each‖
selections are independent and are saved
separately.


MARKer:GROUp {BOTH, EACH }



Marker Chan

Pulse, Mod & Stat

For MK Group both, assigns Marker 1 and

Mk 1 CH

Active, CH1, CH2

Marker 2 to the same measurement channel.

Mk 2 CH

Ext Trig 1, Ext Trig 2

(Mult. Choice)

CH Math,

For MK Group each, assigns Marker 1 and

Mem 1, Mem 2,

Marker 2 to measurement channels.

When Active is selected, the Marker(s) is
assigned to and tracks the active channel.
in the Chan > Selection window.


The Time Mark # assignment for each of the

two group assignments (―Both‖ and ―Each‖)

are independent of each other. When

Fig. 4-29a.

Mark > Exten > Mk Group Both has been Mark > Exten menu

selected, both Marker 1 and Marker 2 are

Both-Active

assigned to the same channel. Changing

either entry changes both.

When Mark > Mk Group Each has been selected, the marker
assignments are recalled from the last selections made in the Mark >
Mk Group Each
mode. Each of the markers can be assigned to separate

measurement channels.

Advertising