ADS Environmental Services Profile Software 950015B6 User Manual

Page 117

Advertising
background image

Profile Main Screen

2-77

CONFIDENTIAL – Covered by ADS Software License Agreement

1998 - 2012 ADS LLC. All Rights Reserved.

Usually Dry Pipe typically remains dry, except under wet
weather conditions.

Electronic Offset Enter the difference that exists between

the depth sensor readings and the manually measured depth.

Deadband Select the distance from the sensor face within

which readings received will be ignored. The default value
(Auto) represents a percentage of the flow depth and should
be sufficient under normal circumstances. However, consider
another value when the sensor readings reveal a potential
problem with false drop-outs. Choose a value for the
deadband that is sufficiently below the minimum depths that
legitimately occur at the monitoring point, yet above the
confirmed drop-outs that have been detected. For example,
choosing 1 inch causes all ultrasonic depth ranges measured
within 1 inch of the face of the sensor to be ignored.

Store Data Select this checkbox to ensure the monitor logs

the ultrasonic data to memory. This option is selected by
default.

Advanced Select this button to access the Advanced

Parameters window.

Note: Modifications made to the Advanced Parameters
should be performed only by or under the guidance of
trained ADS personnel. Therefore, ADS recommends
leaving the default settings in place.

Advanced Parameters

Advertising