entered through the Installation Generator and used when
calculating the depth of the flow.

Channel Select the channel number of the port on top of the

monitor to which the combo sensor is connected.

Smart Depth tab

Surface Combo Device

Electronic Offset Enter the difference that exists between

the smart depth sensor readings and the manually measured
depth. Profile applies this value to the range reading to
calculate depth. ADS recommends obtaining this value during
the confirmation process. Profile defaults to 0 inches.

Mode Select the mode in which you want the smart depth

sensor to operate based on environmental factors or site
constraints. ADS recommends using the High Power
(default) option under most conditions. However, if erratic
Smart Depth readings occur, consider using the Normal
setting.

Store Data Select this checkbox to ensure the monitor logs

the smart depth data to the monitor memory. This checkbox is
selected by default.

Advanced Select this button to access the Advanced

Parameters window.

Note: Modifications made to the Advanced Parameters
should be performed only by or under the guidance of
trained ADS personnel. Therefore, ADS recommends
leaving the default settings in place.

DAC Gain Choose the method for managing the
amplification of the analog signal that returns to the sensor.
Select the Auto (default) option to allow the monitor to
automatically optimize the amplification of the signal.
Select the Manual option to enter a fixed value between -
12dB (lower amplification) and +12dB (higher
amplification).

Surcharge Detection Select the Disable checkbox to
prevent the monitor from implementing the Surcharge