Profile Main Screen

2-45

CONFIDENTIAL – Covered by ADS Software License Agreement

1998 - 2012 ADS LLC. All Rights Reserved.

when the difference in readings falls back below the
designated tolerance.

Enable Depth/Tolerance Select the Enable Depth
checkbox to ensure the monitor logs a CROSS_DEPTH
event when the difference between at least two of the depth
readings at the monitoring point exceeds the corresponding
tolerance. Enter the difference in depth measurements
beyond which you want the monitor to log a cross check
event in the Tolerance field.

Enable Velocity/Tolerance Select the Enable
Velocity checkbox to ensure the monitor logs a
CROSS_VELOCITY event when the difference between at
least two of the velocity readings at the monitoring point
exceeds the corresponding tolerance. Enter the difference
in velocity measurements beyond which you want the
monitor to log a cross check event in the Tolerance field.

Rain Alarm Use the parameters in this section to configure

the rain gauge device for rain alarms. Enter the total amount
of rainfall that must occur within a specified time period to
initiate the alarm in the Threshold field. Then, select the
specific period of time from the Period to Total Rain drop-
down list.

Sampling Use the information in this section to configure

the water quality sampling activities.

Select the type of sampling you want to perform from the
Type drop-down list. Select Depth when you want to
trigger water-quality sampling at a specific flow depth;
select Flow Rate when you want to trigger sampling based
on a specific rate of flow; and select Flow Total when you
want to initiate sampling activities based on a specific flow
volume.

Select or enter the reading at which you want the monitor
to pulse the sampler based on the selected sampling option
and unit of measure in the Trigger field.