the option selected for the Channel in the Device Parameters
section at the bottom of the dialog for that sensor.

Store Data Select this checkbox to ensure the monitor logs

the temperature data to memory.

Note: The remaining parameters (Temperature, Pulse
Command
, and Spare 2 Delay) apply only to Smart
Depth devices associated with ADS 3600, 4000, 5500, and
5600 monitors.

Temperature Select the appropriate temperature sensor for

the monitor to reference when compensating for the air
temperature in the pipe when calculating range.

Pulse Command Enter the strength of the ultrasonic

transmit signal.

Spare 2 Delay Enter the length of time the ultrasonic depth

sensor waits before listening for the return signal (following
the initial transmit signal). ADS does not recommend
changing this parameter unless the data provides erroneous
readings indicating a full pipe that is not actually full. One
Spare 2 Delay unit is equivalent to adding approximately 0.33
inches of distance to the deadband required to receive a
legitimate reading. This should have the same value as the
Spare 2 Delay designated for the ultrasonic device for the
same sensor. Profile defaults to 1.

Surface Combo Device

Edit the Surface Combo 1 (or 2) device associated with a
FlowShark Triton or FlowHawk monitor that represents the
downward-looking ultrasonic range (depth), surface velocity,
surcharge peak velocity, and surcharge pressure depth sensors in the
following way.

The Edit Surface Combo 1 (or 2) Parameters consists of five tabs
for editing the individual sensor parameters containing the surface
combo configuration parameters. Use these parameters to configure
the surface combo sensor device.