6 starting a process – INFICON IQM-233 Thin Film Deposition Controller PCI-Express Card Codeposition Software Operating Manual User Manual

Page 109

Advertising
background image

2 - 89

IP

N 07

4-

58

5-

P1

A

IQS-233 Operating Manual

15b

Click Enabled or Ethernet Enabled to enable RS-232 or Ethernet remote
communication, respectively. Refer to

section 2.3.3.6.6 on page 2-58

.

15c

Click OK to save changes.

16

Click Edit >> System and select the Display tab (refer to

Figure 2-36 on page

2-44

).

17

Select Display and Graph settings to be viewed in the IQS-233 Codeposition
window, refer to

section 2.3.3.6.7 on page 2-60

.

18

Click OK to save changes.

19

Click File >> Save to save the current Process and System parameter values
to the configuration filename displayed in the message area of the
IQS-233 Codeposition window. If the default configuration file, InitConfig.xml, is
displayed, the default parameters loaded when IQS-233 Codeposition software
is started will be overwritten by any changes made to the Process and System
parameters.

19a

Click File >> Save As to save the current Process and System
configuration under a different name. Configuration files are saved in .xml
format. The default folder to save a configuration file to is C: >> INFICON
>> IQS-233 Codeposition
; however, the configuration file may be saved
to another folder location if desired (refer to

Figure 2-11 on page 2-13

).

20

Click Edit >> System and select Card tab (refer to

Figure 2-41 on page 2-56

).

21

Select Enabled in the Simulate pane.

22

Click OK to save changes.

23

Click Start Simulate on the IQS-233 Codeposition window (refer to

Figure 2-56

on page 2-74

) to simulate the process prior to making an actual deposition.

NOTE: Disconnect outputs from the IQM-233 or SQM-242 card(s) to prevent the

supply of power to source outputs.

2.4.6 Starting a Process

1

If in Simulate Mode:

1a

Click Edit >> System and select Card tab (refer to

Figure 2-41 on page

2-56

).

1b

Deselect Enabled in the Simulate pane.

1c

Click OK to save changes.

2

Connect the appropriate inputs and outputs to IQM-233 or SQM-242 card(s).

NOTE: Refer to the IQM-233 or SQM-242 operating manual for information

about the physical input and output connections.

Advertising