INFICON IQM-233 Thin Film Deposition Controller PCI-Express Card Codeposition Software Operating Manual User Manual

Page 75

Advertising
background image

2 - 55

IP

N 07

4-

58

5-

P1

A

IQS-233 Operating Manual

Soak2 Hold . . . . . . . . . . . . . Power stays at the Ramp 2 Power value indefinitely.

Soak2 Release . . . . . . . . . . Deposition phase begins when the Soak 2 Time

value elapses.

PLC Comm pane

The communication settings required by the PLC are configured here.

Port (9600, E, 7, 2) . . . . . . . Disabled, COM#

Select the Comm serial port used to communicate
with the PLC, or select Disabled to disable
communications with the PLC (has same effect as
selecting Not Assigned for all Relay Events and
Input Events). The fixed communication parameters
are: Baud Rate = 9600, Parity = E, Data Bits = 7,
Stop Bits = 2.

Address (CJ1) . . . . . . . . . . . 0 to 100

A single PLC will usually use Address 0. Multiple
PLCs can be controlled from a single computer serial
port by connecting their expansion ports. The unique
slave address of each PLC is usually selected by a
rotary or DIP switch (refer to the PLC operating
manual).

Advertising