2.3.3.5.10 Source/Sensor Tab

Max Power, Slew Rate, and Sensor Tooling are configured in the Source/Sensor
tab, and the material to be deposited is selected in this tab. See

Figure 2-31

.

Figure 2-31 Film Edit window - Source/Sensor tab

Source pane

Material . . . . . . . . . . . . . . . . Select the material to be deposited from a list of

material names. The Density and Z-Ratio values for
that material are displayed in the Materials window
(see

Figure 2-33 on page 2-42

) where these values

can be edited if desired.

Max Power (%) . . . . . . . . . . 0.00 to 100.00%

The maximum Output Power allowed, as a
percentage of the Full Scale Out value in the Outputs
tab on the System Setup window (see

Figure 2-37 on

page 2-45

). For example, if the Full Scale Out value

is -10 and the Max Power value is 75, the Output
Power will not exceed 75%, and therefore the output
voltage to the source will not exceed –7.5 V.

Slew Rate (%) . . . . . . . . . . . 0.0 to 100.0%

The maximum allowed change per second of Output
Power, for an Output using PID loop control, as a
percentage of the Full Scale Out value in the Outputs
tab on the System Setup window.