Binary Coded Pocket Select . . . . . Pockets are selected by a value that is the

binary representation of the pocket, requiring
fewer relays. For example, 16 pockets can
be selected with four relays.

NOTE: Please refer to the indexer operating manual for information regarding

pocket selection.

3.2.2 Other Digital I/O

Depending on the PLC, additional relay and input pins are available for other
commands.

3.3 PLC Setup and Test

1

In IQS-233 Codeposition software click Edit >> System >> I/O.

2

Set the Port (9600, E, 7, 2) to match the serial port connected to the PLC.

3

Set the PLC Comm Address (CJ1) to match the assigned PLC Address
(usually 0). See

Figure 3-1

.

Figure 3-1 System Setup window - I/O tab