2.17 Storing determination results and silo calculations

751 GPD Titrino

99

2.17 Storing determination results and silo calcula-
tions

2.17.1 Storing determination results

If the sample-specific data of the silo memory should be kept after the determination and

supplemented by results, the following entries are necessary:

1. In the method under <DEF>

Assignment of the determination results to C24 and/or C25:

2. In the silo memory, <SMPL DATA> (when the silo memory is switched in):

"save lines: on"

Assignment of determination results

The determination results are assigned in key <DEF>.

The display texts of the Titrino are shown to the left.
The values are the default values.

>silo calculations

C24=
C25=

Silo calculations

Assignment to C24 (RSX, EPX, CXX)

Calculated results (RSX), endpoints (EPX) or variables

CXX can be stored as C24.

Same procedure for C25.

Important:

Ensure that there is still sufficient space for storing the results C24 and C25. (In the report

<PRINT><USER METH><ENTER> the number of free bytes is shown.) Result name,

value and unit are stored. The memory requirements can be estimated as follows:

Result with text (8 characters) and unit (5 characters):

32 bytes

Measured value C40, value without unit:

22 bytes

DEF

(

2

def
>formula

>silo calculations

>common variables
>report
>mean
>temporary variables