2.12 TIP, Titration procedure

751 GPD Titrino

85

2.12 TIP, Titration procedure

TIP (Titration Procedure) is used to link several commands in a sequence.

TIP is selected with <MODE> and <ENTER>. TIP is an "empty shell" in which the

sequence of the determination must be defined.

Definition of the sequence

With key <DEF> the TIP can be defined.

sequence:
of TIP.

The display texts of the Titrino are shown to the left.
The values are the default values.

>sequence

1.step: OFF

<ENTER>

1.method: 5-TIP

<ENTER>

etc

.

2 x

<QUIT>

Sequence

Select a step with keys <

> and <

>:

method:

Method from the user memory or from

the card.

pause:

Waiting time

L4, L6 output: Set an output.

info:

Hold sequence and write a message

into the display.

prep

Preparation of titrating burets.

stirrer

ON/OFF

Confirm the step with <ENTER> and enter the

parameter for the selected step, see below.

The request for the second step follows etc. Up to 30

steps can be selected.

On completion of the sequence definition, exit the

inquiry with <QUIT>.

DEF

(

2

def

>sequence

>formula
>silo calculations
>common variables
>report
>mean