Metrohm 751 GPD Titrino User Manual

Page 41

Advertising
background image

2.6. Parameters, key <PARAM>

751 GPD Titrino

38

req.smpl size: OFF
cond.

limit smpl size: OFF
cond.

low lim. 0.0
cond.
up lim. 999999
cond.

Oven: no
cond.

activate pulse: OFF
cond.

Request of sample size after start of titration (value,

unit, all, OFF)

"all" the value and the unit will be requested.

Limiting value check for sample size (ON, OFF)

With "on" the error message "sample size out." appears

if the entry is outside the set limits. The limiting values

are shown in the display window.

The absolute value of the limit is checked during sam-

ple size input and during the calculation of the results.

If "on" has been set:

Lower limit for sample size (0.0...999 999)

Upper limit for sample size (0.0...999 999)

Connected oven (COM1, COM2, no)

COM of the Titrino to which the oven is connected.

If an oven is connected via RS232 an inquiry will be

made for the oven results and these will be inserted

into the result report of the Titrino. The report output on

the oven must be switched OFF.

Set "no" if no oven has been connected or if you have

not connected the oven to Titrino the via RS232 inter-

face.

Pulse output on I/O line L6 (L6, pin 1) of the remote

socket (first, all, cond., OFF) e.g. start of a Dosimat,

see page 224.

Advertising