6 analogue input test, 1 examination of highly insulated measuring inputs – Metrohm 751 GPD Titrino User Manual

Page 200

Advertising
background image

4.2. Diagnosis

751 GPD Titrino

197

6 Analogue input test

Press <7>.

analog input test 1...5

6.1 Examination of highly insulated measuring inputs

Connect ‘Ind I’ measuring input to a voltage calibrator (e.g. Metrohm 642 pH simulator) by

means of a highly insulated cable (e.g. 6.2108.060). Set calibrator to 0 V.

Press <1>.

Input 1

0.0 mV

Tolerance: ± 0.5 mV

Set the calibrator voltage, on setting 'low ohmic' (with 642 = ~ 0.002 M

) to different val-

ues (e.g. +1500 mV) and compare with the displayed value.

Tolerance (with ± 1500 ÷ 2000 mV) ± 1 mV. Be aware of the calibrator's tolerance.

Set simulator to high ohmic condition (with 642 = 1000 M

).

The displayed reading may vary slightly only (with 1500 mV ó 1 mV)

Press <Clear>.

analog input test 1...5

Disconnect simulator from ‘Ind I’ input and connect to ‘Ind II’ input.

Press <2>.

Input 2

XX mV

Carry out the same measurements as with Input 1.

Press <Clear>.

analog input test 1...5

Short-circuit input ‘Ind I’ (e.g. with cable 3.496.5070).

Press <3>.

Input 1-2

XX mV

The differencial voltage between inputs ‘Ind I‘ and ‘Ind II’ is displayed.

Example: 0 - (+)1500 mV = -1500 mV

Remove cables from the inputs ‘Ind I’ and ‘Ind II’.

Press <Clear>.

analog input test 1...5