Metrohm 751 GPD Titrino User Manual

Page 43

Advertising
background image

2.6. Parameters, key <PARAM>

751 GPD Titrino

40

Control parameters for KFT

The control parameters can be set according to your samples. The default parameters

are already set to get satisfactory results. Optimize the control parameters for specific

samples only.

During the titration, reagent dosing is carried out in 3 phases:

1. Initial dosing:

Here the dosing rate

increases constantly up to

"max.rate".

2. Continuous dosing:

Dosing is performed at the

maximum rate "max.rate"

until the control range

(dynamics) is reached.

3. Control range (dynamics):

In this range, dosing is

performed in single steps.

The last dosing steps are

controlled by the

parameter "min.volume

incr.".

Remarks:

Titer determination should have the same modes and the same parameters as the

sample titrations.

In most cases the preset default parameters provide precise and reproducible results.

Ipol gives relatively steep curves. They are flatter with Upol.

U/mV

V/ml

t/s

EP

V/ml

Initial

dosing

Continuous

dosing

Control

range

I/uA

V/ml

EP

KFT Ipol

KFT Upol

Advertising