.T.B

Begin of sequence (or submethod).

.T.F

Final: End of determination (or submethod), the final steps will be
carried out.

.T.E

Error. Message together with error number, see page 111ff.

.T.H

Hold: Status 'Hold' has been reached.

.T.C

Continue: Continue after hold.

.T.O

Conditioning OK: EP reached (in SET, KFT with conditioning).

.T.N

Conditioning Not OK: EP not reached (in SET, KFT with conditioning).

.T.Re

Request: In the inquiry of an identification or the sample size after
start of titration.

.T.Si

SiloEmpty: Silo empty, i.e. the last line has been removed from the
silo memory.

.T.M

MeasList: Entry in the measuring point list (with DET, MET, STAT,
DOS, DOC, SET, KFT, MEAS).

.T.EP

EPList: Entry into EP list (with DET, MET, SET, KFT)

.T.RC

Results have been recalculated.

Messages from node .C, Comport:
.C.B1

COM1: A report is outputted on COM1. During this time, COM2 will
be blocked. COM2 is generally blocked, if COM1 is busy.

.C.R1

COM1 is ready again. (Comes also when you <QUIT> an error.)

.C.B2, .R2 Identical for COM2.
Messages for changings in the I/O lines. If the changings are made simultane-
ously, there is 1 message. Pulses receive 2 messages: one message each for
line active and inactive.
.I

Input: Change of an input line.

.O

Output: Change of an output line (except 7, pin 2, for recorder
pulses).

If a change occurs that requires a message, the Titrino sends space (ASCII 32)
and ! as an introducer. This is followed by the name of the device (see
3.2.2.106). Special ASCII characters in the device name are ignored. If no de-
vice name has been entered, only ! is sent. Finally the Titrino sends the infor-
mation which node has triggered the message.
Example: !John".T.Si": The message was triggered from instrument "John",
node .T.Si

3.2.2.

3.2.2.170.

170. Se

Setup.G

.Graphics.G

.Grid

ON

ON, OFF

Se

Setup.G

.Graphics.F

.Frame

ON

ON, OFF

Se

Setup.G

.Graphics.S

.Scale

Full

Full, Auto

Se

Setup.G

.Graphics.R

.Recorder.R

.Right

0.2...0.5

0.5...1.00

Se

Setup.G

.Graphics.R

.Recorder.F

.Feed

0.01...0.05

0.05...1.00

Change in the appearance and the format of the curve for the output. The set-
tings are valid for both Titrino COM ports.
.Grid:

On/off switching of grid over curve.

.Frame:

On/off switching of frame surrounding the curve. If grid and frame
are switched off, the curve is printed faster as the print head does
not have to move to the end of the paper.

.Scale:

Type of scaling of the measured value axis: Full means that the
scale runs from the smallest up to the greatest measured point.
With auto, the smallest measured value is taken and the next
smaller tick defines the beginning of the scale; the next greater tick
to the greatest measured value is the end of the scale.