Metrohm 751 GPD Titrino User Manual

Page 26

Advertising
background image

2.6 Parameters, key <PARAM>

751 GPD Titrino

23

EP recognition criteria for DET and MET
The parameter "EP recognition" offers you a range of possibilities to ensure selection of

the EP you are interested in: If the desired jump is very large, you can select the "greatest"

jump (with DET the steepest jump will be evaluated). Thus you always obtain just one EP

per titration (EP1).

If you wish to determine the sum of different components (e.g. acid or base numbers), the

"last" jump can be the correct one.

And finally you can set a "window" for each expected EP.

EP windows

EP windows are used

to suppress disturbing influences and EP's which are not needed.

to increase the liability for the calculation of the results. The EP windows make an

unequivocal assignment of the EP's possible: per window one EP is recognized; the

numbering of EP's is defined by the windows so that even if EP's are missing, the

calculations are still performed with the correctly assigned EP volumes.

An EP window defines the range in which an EP is expected. EP's outside these ranges

are not recognized. Windows are defined on the measured value axis.

2 EP's are recognized. Their

numbering is defined by the windows:

Window 1

EP1

Window 2

EP2

If more than 1 EP is expected, a

window must be set for each EP.

Windows must not overlap. They may

only touch each another.

Rule: If there are more than 1 EP in a

window, the first jump is recognized as

EP1, the second is not recognized.

EP1 is marked as EP1 + to indicate

that more than one EP has been found

in the window

U/mV

V/ml

EP window 2

EP window 1

EP 1

EP 2

U/mV

V/ml

EP 1

+

EP window 1

lower lim. 1

upper lim. 1