Metrohm 751 GPD Titrino User Manual

Page 63

Advertising
background image

2.6. Parameters, key <PARAM>

751 GPD Titrino

60

Filling times
Refilling times are not incorporated by the Titrino in the calculation of the rate. The filling

times can be calculated using the following formula:

max. filling rate

filling rate in s = ДДДДДДДДДДДДДДД

*

20 s + 3 s

current filling rate

The max. filling rate depends on the Exchange unit installed, see page 56. The filling

times (incl. stopcock rotation) can be estimated with the aid of the following table and

taken into account in your dosing routine:

Exchange

Unit

Filling rate

"max."

Filling rate

"100 ml/min"

Filling rate

"50 ml/min"

Filling rate

"10 ml/min"

5 ml

23 s

---

---

33 s

10 ml

23 s

---

---

63 s

20 ml

23 s

---

27 s

123 s

50 ml

23 s

33 s

63 s

303 s

Example:
Dispensing of 1 L reagent is required over a period of 1 hour with a 50 mL Exchange unit.
Calculate the rate to be set.

Dispensing time = 60 min.
Dispensed volume = 1000 mL
Filling time = 23 s (filling rate = max.)
Volume of the Exchange Unit = V(B) = 50 mL

Number of refills = Dispensed volume/V(B) = 1000/50 = 20

If this division does not result in a remainder, the last filling is no longer a "refilling" and
must be deducted. In our example, 19 refills are required. During this time, dispensing is
not performed, i.e. the actual dispensing time is 3600 s - 19 * 23 s = 3163 s =
52.717 min.

The dispensing rate is thus given by 1000 mL/52.717 min = 18.972 mL/min

Summary in a formula:

Volume

1000

Rate = ДДДДДДДДДДДДДДДДДДДДДДДДДДДД = ДДДДДДДДДДДДДД=18.972 ml/min

Disp.time - Refills * Fill.time * 1/60

60 - 19 * 23 * 1/60