Metrohm 751 GPD Titrino User Manual

Page 62

Advertising
background image

2.6 Parameters, key <PARAM>

751 GPD Titrino

59

Sequence with DOS

After the start, the activate pulse is output.

The start delay time is waited off.

The sample identifications and the sample size are re-
quested.

The pause is waited off. During this time the limits are not
yet monitored.

During the dosing, measured values (time, volume) are
acquired and stored in the measuring point list. If moni-
toring of measured values and/or temperature is active,
these values will also be stored in the measuring point list.
If limits are violated it will be marked in the measuring
point list.
If a T sensor is connected, the temperature is measured
continuously (in intervals of 2 s). The pH will be corrected
as a function of temperature.
Without T sensor the values are measured without inter-
rupt by T measurements. The manually set temperature is
then valid.
In dosings which use "time" or "rate" as dispensing type,
the volume goes up to 99 999.99 ml. After that it will be
set to zero and the dosing continues.
The dosing is terminated when the dosing volume or the
dosing time is reached.

Calculations are carried out.

Data are output. The measuring point list is the base for
curves.

(Activate pulse)

<START>

(Start delay)

(Pause)

(Request ident.)

(Request smpl size)

Dosing

Acquisition of meas.

(Monitoring)

Calculations

Data output

Advertising