4.2. Diagnosis

751 GPD Titrino

198

6.2 To check Pt 100 / 1000 connection

Connect a Pt 100 or Pt 1000 sensor, a resistor switch-box or a single resistor of 100

or

1 k

, respectively, to sockets ‘Pt 100/1000’ by means of short cables.

Press <4>.

Pt 100*

XX °C

(* or Pt 1000)

Tolerance: ± 0.5 °C (Note also tolerance of resistor switch-box).

The sensor allows automatic displaying of the room temperature. (The resistor values

correspond to 0°C.)

Press <Clear>.

analog input test 1...5

Remove cables and resistor switch-box.

6.3 Polarizer test

Press <5>.

polarizer test

Press <Enter>.

dummy resistor 14.3kΩ

Ω ?

Connect resistor switch-box (or suitable resistor 14.3 k

0.1%) using 3.496.5070 cable to

‘Pol’ socket. Switch-box to 14.3 k

.

Press <Enter>.

Test sequence:

1. An asterisk flashes during the test.
2. In case of an error an error message appears. (If for example the switch-box is not

connected,

error 100 appears).

3. If no error is found, after about 15 s display shows

polarizer test o.k.

Press <Clear>.

analog input test 1...5

Press <Clear>.

diagnose press key 0...9

Remove cable and resistor switch-box.