Metrohm 751 GPD Titrino User Manual

Page 12

Advertising
background image

2.4 Configuration, key <CONFIG>

751 GPD Titrino

9

stirrer control: OFF

remote box: OFF

keyboard: US

barcode: input

Automatic switching ON/OFF of the stirrer in the titration

sequence (ON, OFF)

If stirrer control is ON, the stirrer will be switched on at
the beginning and switched OFF at the end of a de-
termination. For SET and KFT with conditioning the
stirrer will be switched off in the inactive state.
In the modes MEAS, CAL, DOS, and TIP the stirrer is
not switched automatically.
For stirrer control the red switch on the stirrer unit must
be ON.

Connection of a remote box (on ,OFF)

To the remote socket for PC keyboard and barcode

reader, see page 212.

If "on" has been set:

Type of PC keyboard (US, German, French, Spanish,

Swiss.)

The PC keyboard is used as an input aid, see page

213.

Target for barcode reader (input, method, id1, id2, id3,

smpl size)

The barcode reader is used as an input aid, see page

212.

Input:

The barcode string goes to the entry field in

which the cursor is currently located.

Method:

The barcode string goes to the entry field

"Methods" in the silo memory.

Id1:

The barcode string goes to the entry field

"Id1". (Similar for Id2 and Id3.)

Smpl size:

The barcode string goes to the entry

field "smpl size".

>auxiliaries

dialog: english

date 1998-04-23

time 08:13

run number 0

auto start OFF

General settings

Selection of dialog language (english, deutsch,

francais, español, italiano, portugese, svenska)

Current date (YYYY-MM-DD)

Format: Year-month-day, entry with leading zeros.

Current time (HH-MM)

Format: Hours-minutes, entry with leading zeros.

Current run number for result output (0...9999)

The sample number is set to 0 when the instrument is

switched on and incremented on every determination.

Automatic starts of titrations. (1...9999, OFF)

Number of automatic starts ("number of samples").

Used for instrument interconnections in which the

external instrument does not initiate a start. Not

advisable in connections with Sample Changers.

Advertising