2.12. TIP, Titration procedure

751 GPD Titrino

86

Information for the commands (steps):

Command

Meaning

Input range

method

Method from the user memory or from the card. This

method runs as a submethod.

Name

pause

Waiting time. The waiting time can be aborted with

<QUIT>.

<CLEAR> sets "inf" (= infinitely long pause time).

0...999 999 s,

inf.

L4, L6 output

Set L4 output (pin 3) resp. L6 output (pin 1) of the

remote socket.

active = 0 V, inactive = 5 V, pulse > 100 ms, off =

output is not used.

Cable Titrino (L6) - Dosimat: 6.2139.000.
Important: A pulse (e.g. a pulse from monitoring or an

activate pulse in a submethod) can set an active output

to inactive! At the end of the TIP method, the outputs

are set to "inactive".

active,

inactive,

pulse, off

info

Message in the display .

The TIP sequence is held and the message displayed.

Continue the sequence with <START>, <QUIT> or

<ENTER>.

up to 16

characters

prep

Preparation of titrating burets.

internal D0,

external D1,

external D2

stirrer

Switching stirrer ON/OFF.

In TIP the stirrer is not switched automatically in the

submethods. At the end of TIP the stirrer is switched off

(if stirrer control is ON).

ON, OFF

The parameters of the sequence can be viewed and changed at any time under the
<PARAM> key.