7 result calculations – Metrohm 751 GPD Titrino User Manual

Page 77

Advertising
background image

2.7. Result calculations

751 GPD Titrino

74

2.7 Result calculations

Formula entry, key <DEF>

Key <DEF> contains various inquiries for result

calculations and data output. The data of this key are

method specific and they are stored in the method

memory together with the method.

formula:
Formulas for result calculations.

The display texts of the Titrino are shown to the left.
The values are the default values.

>formula

RS?

RS1=

RS1=EP1*C01/C00

Input of formulas

Enter formula number (1...9)

You can calculate up to 9 results per method.

Enter a number 1...9.

Input of formula

Example:

RS1=EP1*C01/C00

Enter formula by means of 3rd functions of keyboard.

Here you will find operands, mathematical operations

and parentheses. Operands require a number as an

identification. You can use the following operands:

EPX: EP's. X = 1...9

RSX: Results which have already been calculated with

previous formulas. X = 1...9.

CXX: Calculation constants. XX = 00...89.

Rules:

Calculation operations are performed in the

algebraic hierarchy: * and / before + and -.

Store formula with <ENTER>.

Calculation quantities and operands can be deleted

with <CLEAR> one by one.

To delete a complete formula press <CLEAR>

repeatedly until only RSX remains in the display.

Confirm with <ENTER>.

If a formula is stored with <ENTER>, result text,

number of decimals and result unit will be requested:

DEF

(

2

def

>formula

>silo calculations
>common variables
>report
>mean
>temporary variables

Advertising