Advertising
background image

3.2 Remote control commands

751 GPD Titrino

153

wait:

Interrupt dosing until the limit value is no longer violated, then con-
tinue automatically.

3.2.2.

3.2.2.50.

50. M

Mode.P

.Parameter.M

.Monitoring.L

.L10Output.A

.AssignOutput

meas, temp, rate, all, none

none

M

Mode.P

.Parameter.M

.Monitoring.L

.L10Output.L

.Limit

upper,lower,any

any

M

Mode.P

.Parameter.M

.Monitoring.L

.L10Output.O

.Output

active, pulse

pulse

Parameters for STAT, DOS, DOC: Output of a signal on L10 output (pin 8) of
the remote socket on limit value violation.
The output can be assigned a definite or all monitoring value(s) as well as the
violated limit and the type of signal to be outputted determined.

3.2.2.

3.2.2.51.

51. M

Mode.P

.Parameter.P

.Presel.R

.RateDisp

ON, OFF

OFF

Parameters for STAT, DOC: Display of the current rate during the determination.

3.2.2.

3.2.2.52.

52. M

Mode.P

.Parameter.D

.DosPara.T

.Type

volume

volume, time, rate

M

Mode.P

.Parameter.D

.DosPara.V

.Volume.V

.Volume 0.000...10

10...99999.99

M

Mode.P

.Parameter.D

.DosPara.V

.Volume.D

.DisType

time, rate

rate

M

Mode.P

.Parameter.D

.DosPara.V

.Volume.R

.Rate

0.001...150, max.

max.

M

Mode.P

.Parameter.D

.DosPara.V

.Volume.T

.Time

1...100

100...999 999

M

Mode.P

.Parameter.D

.DosPara.Ti

.Time.T

.Time

1...100

100...999 999

M

Mode.P

.Parameter.D

.DosPara.Ti

.Time.D

.DisType

volume

volume, rate

M

Mode.P

.Parameter.D

.DosPara.Ti

.Time.R

.Rate

0.001...150, max.

max.

M

Mode.P

.Parameter.D

.DosPara.Ti

.Time.V

.Volume

0.000...10

10...99999.99

M

Mode.P

.Parameter.D

.DosPara.R

.Rate.R

.Rate

0.001...150, max.

max.

M

Mode.P

.Parameter.D

.DosPara.R

.Rate.S

.StopType

volume

volume, time

M

Mode.P

.Parameter.D

.DosPara.R

.Rate.T

.Time

1...100

100...999 999

M

Mode.P

.Parameter.D

.DosPara.R

.Rate.V

.Volume

0.000...10

10...99999.99

Parameters for DOS: In the equation Rate = Volume/Time, two parameters are
preset, the third is calculated. Depending on the entry of the type, the following
information is required: .Volume, .Time or .Rate.
Entries: .Volume in ml, .Time in s, .Rate in ml/min.

3.2.2.

3.2.2.53.

53. M

Mode.P

.Parameter.D

.DosPara.P

.Pause

00...999 999

Parameters for DOS: Pause time in s.

3.2.2.

3.2.2.54.

54. M

Mode.P

.Parameter.D

.DosPara.TD

.TDelta

1...10

10...999 999

Parameters for DOS, DOC: Time interval for the entry of a measured point in the
list of measured points.

3.2.2.

3.2.2.55.

55. M

Mode.P

.Parameter.D

.DosPara.Te

.Temp

-170.0...25.0

25.0...500.0

Parameters for DOS, DOC: Temperature in °C. The temperature is measured
continuously and updated if a T sensor is attached.
The temperature is used for temperature compensation in pH measurements.