Metrohm 751 GPD Titrino User Manual

Page 81

Advertising
background image

2.8. Statistics calculation

751 GPD Titrino

78

How do you obtain statistics calculations?
1) Enter the allocations for the statistics calculation, see page 77.

2) Switch on the statistics calculations: Either with <STATISTICS> or set the status under

<PARAM>, "> statistics" to "ON". The "STATISTICS" LED is on. Storing a method in

the method memory, the status of the statistics calculation is retained.

3) Change the number of the individual values n under "mean n", if necessary.

4) Perform at least 2 titrations. The statistics calculation are constantly updated and

printed. The values are printed in the short and full result report.

5) The statistics report can be printed with <PRINT><STATISTICS><ENTER>.

Rules:

Recalculated results are incorporated in the statistics calculation.

If a result of a particular titration can not be calculated, no results for this determination

are incorporated in the statistics calculation. However, the sample counter is still

operative, i.e. the statistics calculation start afresh when the number of required

individual determinations has been performed.

If the statistics are switched off ("statistics" LED no longer on), results are no longer

entered in the statistics table. But the table remains unchanged. When the statistics

are switched on again, you can immediately continue working.

If you delete results, all results of the determination with index n are removed from the

statistics evaluation.

On method change, the old statistics table is cleared and the statistics instructions of

the new method followed.

Old results in the statistics table can be deleted with "delete all" (<PARAM>,

">statistics", "res.tab:").

Advertising