10 internal dosing unit d 0 – Metrohm 751 GPD Titrino User Manual

Page 204

Advertising
background image

4.2. Diagnosis

751 GPD Titrino

201

9 Memory card test

Press <9>.

memory card test

Insert memory card.

The memory card has to be formated from a 751 GPD Titrino otherwise the test does not recognize the card

and on the display appears „memory card 000 KB end“.

Press <Enter>.

The test runs automatically. If no fault arises then there appears:

memory card 128 KB o.k.

Press <Clear>.

diagnose press key 0...9

10 Internal dosing unit D 0

Insert exchange unit and put the burette tip into a collecting receptacle.

Press <Clear>.

The Titrino gets out of the diagnosis menu and returns to the user program.

The spindle drive and the cock changeover drive of the internal dispensing unit can be
checked with the aid of the function <prep> (see page 99).

Set the parameters of this function accord. to the following list (for setting the parameters

see page 9).

configuration

>prep.dosing elements

power ON prep: OFF

report: OFF

dos.element: internal D0

warn.interv.D0 OFF min

volume D0 3.5 ml

cycles D0 2

dos.rate DX max. ml/min

fill rate DX max. ml/min

Press <Prep>.

intern DO prep

Advertising