4.2. Diagnosis

751 GPD Titrino

199

7 External inputs and outputs

This test is meaningful only if the 751 GDP Titrino is used interconnected with other instruments via the 'Re-

mote' connection. In addition, a 3.496.8550 test plug normally used in the repair service is required for this test.

However, this plug can also be purchased by customers under the above number.

For the sake of completeness, the procedure is described here. If a diagnostic test of the external inputs and

outputs is not required, continue with point 8.

5 18 4 17 3 16 1

2

6

7

8 13 19 20

12 24 11 23 10 22 9 21

Fig. 4

Connections in the 3.496.8550 test plug

Press <4>

extern input/output test

Press <Enter>.

I/O-test-connector?

Insert the 3.496.8550 test plug in port B 'Remote'. (Do not switch off instrument!).

Press <Enter>.

Test sequence:

1. In case of an error an error message is displayed. If for example no test plug is

connected, error message

error 50 01HEX appears).

2. If no error is found, after about 1 s display shows

extern input/output o.k.

Remove test plug.

Press <Clear>.

diagnose press key 0...9

8 RS 232 test

A 3.496.8560 test plug normally used in the repair service is required for this test. However, this plug can also

be purchased by customers under the above number.

For the sake of completeness, the procedure is described here. If a diagnostic test of the RS 232 interface is

not required, continue with point 9.