2.6. Parameters, key <PARAM>

751 GPD Titrino

46

>stop conditions

stop time: OFF
**titr.

stop time 999999 s
t(delta) 999999 s
t(delay) 999999 s
**titr.

factor 999999
**titr.

stop V: abs.
**titr.

stop V 99.99 ml
**titr.

factor 999999
**titr.

stop rate OFF ml/min
**titr.

filling rate max. ml/min
**titr.

Stop conditions

The condition which is met first will be effective.

Type of stop time (abs.,rel., delta, delay, OFF)

"abs": absolute stop time in s

"rel.":

relative stop time to sample size.

"delta" Time after having reached the EP the first time.

"delay" Time after last dosing increment.

"OFF": stop time switched off. Stop time is not

monitored.

See page 52.

If "abs.", "delta" or "delay" are set:

Request of time (0...999 999 s)

If "rel." is set:

Factor for relative stop time (0...±999 999)

Calculated as:

Stop time in s = factor * sample size

Type of stop volume (abs., rel., OFF)

"abs.": absolute stop volume in mL.

"rel.":

relative stop volume to sample size.

"OFF": stop volume switched off. Stop volume is not

monitored.

If "abs." is set:

Absolute stop volume (0...9999.99 mL)

If "rel." is set:

Factor for relative stop volume (0...±999 999)

Calculated as: Stop V in mL = factor * sample size

Stop when dosing rate is smaller than the preset rate.

(0.01...150 mL/min, OFF)

<CLEAR> sets "OFF".

Stop if the current rate is below the stop rate. This

criterion becomes effective 10 s after start.

Filling rate (0.01...150 mL/min, max.)

<CLEAR> sets "max.".

The maximum rate depends on the Exchange Unit:

Exchange Unit

max.

5 mL

15 mL/min

10 mL

30 mL/min

20 mL

60 mL/min

50 mL

150 mL/min