Metrohm 751 GPD Titrino User Manual

Page 37

Advertising
background image

2.6. Parameters, key <PARAM>

751 GPD Titrino

34

2.6.3 Parameters for KFT

control parameters:
for the EP.
titration parameters
control the general course of the titration.
stop conditions:
Parameters for the termination of the titration.
statistics:
Calculation of mean values and standard deviation,
see page 77.
preselections:
ON/OFF of various auxiliary functions such as auto-
matic requests after the start and activate pulse.

>control parameters

EP at U 250 mV
**titr.

dynamics 100 mV
**titr.

max.rate max. ml/min
**titr.

min.volume incr. min. ul
**titr.

Control parameters

Endpoint (input range depending on the measured

quantity:

U, Ipol:

0.. ±2000 mV

I, Upol:

0...±200.0 uA)

Control (input range depends on the measured

quantity:

U, Ipol:

1...2000 mV

I, Upol:

0.1...200.0 uA)

Outside of the control range: constant dosing, see

page 40.

Maximum dosing rate (0.01...150 mL/min, max.)

<CLEAR> sets "max.".

This parameter determines primarily the addition rate

outside the control range, see page 40.

The maximum rate depends on the Exchange Unit:

Exchange Unit

max.

5 mL

15 mL/min

10 mL

30 mL/min

20 mL

60 mL/min

50 mL

150 mL/min

Minimum volume increment (0.1...9.9 ul)

This parameter determines the addition rate at the start

and the end of the titration, see also page 40.

This parameter influences the titration speed and

therefore its accuracy very strongly: A smaller

"min.volume incr." results in a slower titration.

parameters

>control parameters

>titration parameters
>stop conditions
>statistics
>preselections