2.6. Parameters, key <PARAM>

751 GPD Titrino

22

Reagent feeding and EP evaluation of MET
In monotonic titrations, the volume increment is constant over the whole titration curve.

V step

Volume increment.

A prerequisite for good accuracy is the correct size of

the volume increments. A good value is given by V step

= 1/20 V

EP

(V

EP

= volume of the EP). In any case, the

increment size should always be between 1/10 V

EP

and

1/100 V

EP

; with steep jumps preferably in the region of

1/100 and with flat jumps preferably in the region of

1/10. The accuracy of the evaluation can not be

increased by dispensing small increments as the

changes in the measured value can then be of the same

order of magnitude as the noise. This can produce

"ghost EP's"!

EPC

The EP's are localized with an algorithm which is based

on Fortuin and has been adapted by METROHM for

numeric procedures. (METROHM Bulletin 2, No. 10,

1971). Here, the greatest change in the measured value

is sought (

n

). The exact equivalence point is

determined with an interpolation factor, which depends

on the delta values before and after

n

:

V

EP

= V

0

+

ρ

V

V

EP

:

EP volume

V

0

:

Total dispensed volume before

n

V:

Volume increment

ρ

:

Interpolation factor (Fortuin)

Equivalence Point Criteria.

The preset EPC is compared to the found ERC

(Equivalence point Recognition Criteria) for each

evaluated equivalence point. The size of the ERC is

displayed in the titration curve, see page 83. ERC is the

sum of the measured value changes before and after the

break:

∆

n-2

+

∆

n-1

+

∆

n

+

∆

n+1

+

∆

n+2

(In certain cases there are only 3 or 1 summand).

EP's whose ERC is below the preset EPC will not be

recognized. For most titrations the standard value is

suitable.

The evaluation can be repeated at any time after the

titration in a "dry run" with changed evaluation criteria.

The old titration data are not deleted until a new titration

is started.