2.4 Configuration, key <CONFIG>
751 GPD Titrino
13
>curve
grid: ON
frame: ON
scaling: Full
width 0.75
length 0.1
curve
The settings are valid for COM1 and COM2.
If you change the printer type, the following settings are
initialized according to the printer.
Grid drawing (ON, OFF)
Frame drawing (ON, OFF)
Type of scaling (Full, Auto)
Full: The scaling goes from the greatest to the smallest
value.
auto: The scaling from tick to tick, e.g. the
smallest/greatest values lie in between the first/last tick.
Width (0.2...1.00)
1 is greatest width. If you set 1 you may loose the label
at the right margin.
Length (0.01...1.00)
Volume axis
Curve length per cylinder volume
0.01
100 cm
0.1
10 cm
0.5
2 cm
1
1 cm
Time axis:
Curve length
0.01
100 cm
0.1
10 cm
0.5
2 cm
1
1 cm