2.6. Parameters, key <PARAM>

751 GPD Titrino

70

temperature 25.0 °C

time interval 2 s

Temperature (-170.0...500.0 °C)

Temperature is measured at the start if a T sensor is

connected.

This parameter is used for temperature compensation

in pH measurements.

Time interval (1...999 999 s)

Time interval for acquisition of measured values.

>preselections

req.ident: OFF

req.smpl size: OFF

limit smpl size: OFF

low lim. 0.0
up lim. 999999

activate pulse: OFF

Preselections for the sequence

Request of sample identifications after start of titration

(id1, id1&2, all, OFF)

After start, sample identifications can be inquired

automatically: Only id1, id1 & id2, all three id's or no

inquiries.

Request of sample size after start of titration (value, unit,

all, OFF)

Limiting value check for sample size (ON, OFF)

With "on" the error message "sample size out." appears

if the entry is outside the set limits. The limiting values

are shown in the display window.

The absolute value of the limit is checked during sam-

ple size input and during the calculation of the results.

If "on" has been set:

Lower limit for sample size (0.0...999 999)

Upper limit for sample size (0.0...999 999)

Pulse output on line "activate" (L6, pin 1) of the remote

socket (ON, OFF)

see page 223.