3.2 Remote control commands

751 GPD Titrino

151

If an Oven is connected, its results will be incorporated into the result report of
the Titrino.
If there is no Oven connected via RS232, this parameter has to be on "no".

3.2.2.

3.2.2.40.

40. M

Mode.P

.Parameter.C

.CtrlPara.Mi

.MinRate

0.01...25.0

25.0...9999

Parameter for STAT: Control parameters.
.MinRate: Minimum titration rate in ul/min.

3.2.2.

3.2.2.41.

41. M

Mode.P

.Parameter.T

.TitrPara.T

.TStart

00...999 999

M

Mode.P

.Parameter.T

.TitrPara.StartM

.StartMeas

pH: 0.00...

±

20.00, OFF

OFF

U, Ipol: 0...

±

2000, OFF

OFF

Upol: 0.0...

±

200.0, OFF

OFF

M

Mode.P

.Parameter.T

.TitrPara.R

.RateStart

0.01...150, OFF

OFF

Parameters for STAT: Start of the measured point acquisition after the start
condition has been met.
.TStart: Start time in s.
.StartMeas: Start measured value.
.RateStart: Start rate in mL/min. The current rate must be less than the start

rate .This criterion does not become effective until 10 s after the
start.

3.2.2.

3.2.2.42.

42. M

Mode.P

.Parameter.St

.StopCond.T

.TStop.T

.Type abs.,rel.,delta,delay,OFF

,OFF

M

Mode.P

.Parameter.St

.StopCond.T

.TStop.T

.Time

0...999 999

999 999

M

Mode.P

.Parameter.St

.StopCond.T

.TStop.F

.Factor

0...

±

999 99

999 99

M

Mode.P

.Parameter.St

.StopCond.T

.TStop.D

.Delta

0...999 999

999 999

M

Mode.P

.Parameter.St

.StopCond.T

.TStop.Dela

.Delay

0...999 999

999 999

Parameters for STAT: Stop time.
If an absolute stop time (abs.) has been selected, the time is in s (Time).
A relative stop time (rel.) is calculated as a function of the sample size:
Stop time in s = sample size * factor
The factor selected (Factor) applies.
delta: Time in s after control point is reached for the first time.
delay: Time in s after last dosing step.
OFF means that the criterion is not monitored.

3.2.2.

3.2.2.43.

43. M

Mode.P

.Parameter.St

.StopCond.R

.RateStop

0.01...150, OFF

OFF

Parameters for STAT: Stop rate in ml/min. The current rate must be less than
the start rate. This criterion does not become effective until 10 s after the start.

3.2.2.

3.2.2.44.

44. M

Mode.P

.Parameter.E

.Evaluation.T

.TimeWin.1.L

.1.LowLim 0...999 999,OFF

OFF

M

Mode.P

.Parameter.E

.Evaluation.T

.TimeWin.1.U

.1.UpLim 0...999 999, OFF

OFF

Parameters for STAT: Evaluation of rates in time windows. Entry of the times in
s. The evaluated rates are available as variables C8X (X = 1...9).

3.2.2.

3.2.2.45.

45. M

Mode.P

.Parameter.E

.Evaluation.F

.FixVol.1.V

.1.Value

0...999 999, OFF

OFF