2.6 Parameters, key <PARAM>

751 GPD Titrino

47

>evaluation

low lim.1 OFF s
up lim.1 OFF s

fix V1 OFF s

fix time 1 OFF V(tot)

Evaluation

see also page 54.

Evaluation of dosing rates within time windows

(0...999 999 s, OFF)

Up to 9 time windows in which a rate should be

calculated. The rates are available for calculations as

C8X.

For the calculation of a rate, the measuring point list in

the time window must contain at least 4 points. If no

point falls on the window limits, the next outer one is

used.

Interpolation of the volume at a fixed time (0...999 999 s,

OFF)

Up to 9 fix V. The associated volumes are available for

calculations as C5X.

Interpolation of times at volume ratios at the end V (0.01

... 1.00, OFF)

The volume is entered as a fraction of the final volume,

e.g. 0.25 = 25% of the final volume.

Up to 9 fix times. The associated times are available for

calculations as C6X.

>monitoring

meas.val: OFF
**titr.

low lim. pH -20.00
up lim. pH 20.00
**titr

action: none
**titr.

Monitoring of limit values

Out of limits are marked in the measuring point list.

Measured values and temperatures are only stored in

the measuring point list if their monitoring is active.

Monitoring of measured values (ON, OFF)

With "on" follow the requests:

Limits for measured values (input range depends on the

measured quantity):

pH:

0...±20.00

U, Ipol:

0...±2000 mV

Upol:

0...±200.0 uA)

Action if a limit is exceeded (end, hold, wait, none)

end: Abort.

hold: Hold reagent addition until manually restarted.

wait: Hold reagent addition until limits are again

complied with, then continue automatically.