Metrohm 751 GPD Titrino User Manual

Page 30

Advertising
background image

2.6 Parameters, key <PARAM>

751 GPD Titrino

27

dos.rate max. ml/min
**titr.

pause 2 0 s
**titr.

extr.time 0 s
**titr.

dos.element: internal D0

meas.input: 1

I(pol) 1 uA

U(pol) 400 mV

electrode test: OFF

temperature 25.0 °C
cond.

time interval 2 s
cond.

Dosing rate for start volume

(0.01...150 mL/min, max.)

<CLEAR> sets "max.".

The maximum rate depends on the Exchange Unit:

Exchange Unit

max.

5 mL

15 mL/min

10 mL

30 mL/min

20 mL

60 mL/min

50 mL

150 mL/min

Pause 2 (0...999 999 s)

Waiting time after start volume, e.g. reaction time after

dosing of a start volume. The waiting time can be

aborted with <QUIT>.

Extraction time (0...999 999 s)

During this time the titration is already running. It will be

not stopped (also if the EP is already reached) before

the extraction time is terminated. The extraction time

can be aborted with <QUIT>.

Selection of dosing element (internal D0, external D1,

external D2)

internal D0: internal buret of the Titrino

external D1/2: buret D1, D2, resp.

Measuring input (1, 2, diff.)

Inquiry only with measured quantities pH and U.

Measuring input 1 or 2 or differential amplifier;

connection of electrodes, see page 215.

With polarized electrodes, instead of the measuring

input the

polarization current (-127...127 uA)

or the

polarization potential (-1270...1270 mV, in steps of 10

mV)

is inquired.

Electrode test (OFF, ON)

Test for polarized electrodes. Performed on

changeover from the inactive state to a measurement.

"OFF" means that the test is not performed.

Titration temperature (-170.0...500.0°C).

Temperature is measured at the start of the titration if a

T sensor is connected. The value is used for

temperature compensation in pH titrations.

Time interval (1...999 999 s)

Time interval for acquisition of a measured value into

the measuring point list.