4 error messages, troubleshooting, 1 error and special messages – Metrohm 751 GPD Titrino User Manual

Page 190

Advertising
background image

4.1 Error and special messages

751 GPD Titrino

187

4 Error messages, Troubleshooting

Data transfer inoperative See measures on page 185.

4.1 Error and special messages

XXX bytes missing

XXX bytes missing. For the storage of a method or a silo line

XXX bytes are missing or there is insufficient RAM for a TIP

sequence.

Remedy: <QUIT>. Delete methods no longer needed or use

fewer silo lines.

calibrate electrode

The calibration interval is expired.

Exit: Carry out a calibration with mode CAL or delete

calibration by entering data for pHas or slope in key

<CAL.DATA>.

card battery low

The battery tension is between 2.37...2.64 V.

Remedy: <CLEAR>

and change battery as soon as possible.

card busy

The memory card is busy.

card not accessible

The card is not accessible because it is used by the RS

interface or a card report is prepared.

card read/write error

Working with the card a read/write error appeared.

Remedy: <CLEAR> or insert a different card.

card unformatted

Remedy: <CLEAR> and format the card.

card write-protected

Writing functions are not possible, if the card is write-

protected.

Remedy: <QUIT> and remove write protection, i.e. push write

protection to the left.

change card battery

The date for changing the battery is expired. Leave the card in

Titrino during battery change to ensure constant power supply.

Remedy: <CLEAR>, change the card battery and enter a new

battery change date.

check drive unit!

The buret is not connected (correctly) or defect.

Remedy: Rectify fault or <STOP>.

check electrode

With polarized electrodes. There is a break or short circuit.

Possible causes and rectification of the fault:

- the electrode is not plugged in

plug it in

- the electrode is not immersed in the solution

immerse it

- the electrode is defective

use new electrode.

- the electrode cable is defective

use new cable.

The electrode test can be switched off under the <PARAM>

key.

Exit: Rectify fault or <STOP>.

check exchange unit

The Exchange Unit is not mounted (properly).

Exit: Mount Exchange Unit (properly) so that the coupling

engages or <STOP>.

check remote box

The Remote Box is not (correctly) connected or the Remote
Box is connected but not activated under the <CONFIG> key.
Exit: Connect Remote Box (correctly) and set "Remote Box:
ON" under <CONFIG>, >peripheral units. Switch the Titrino
off/on.