Metrohm 751 GPD Titrino User Manual

Page 157

Advertising
background image

3.2. Remote control commands

751 GPD Titrino

154

3.2.2.

3.2.2.56.

56. M

Mode.P

.Parameter.M

.Monitoring.M

.MeasVal.M

.MeasModepH,U,Ipol,Upol,OFF

OFF

M

Mode.P

.Parameter.M

.Monitoring.M

.MeasVal.MeasI

.MeasInput

11, 2, diff.

M

Mode.P

.Parameter.M

.Monitoring.M

.MeasVal.P

.PCurrent

0...11...

±

127

M

Mode.P

.Parameter.M

.Monitoring.M

.MeasVal.PV

.PVoltage

0...400

400...

±

1270

M

Mode.P

.Parameter.M

.Monitoring.M

.MeasVal.Po

.PolElectrTest

ON, OFF

OFF

M

Mode.P

.Parameter.M

.Monitoring.M

.MeasVal.pH.L

.pH.LowLim

00...

±

20.00

M

Mode.P

.Parameter.M

.Monitoring.M

.MeasVal.pH.U

.pH.UpLim

0...

±

20.00

20.00

M

Mode.P

.Parameter.M

.Monitoring.M

.MeasVal.U.L

.U.LowLim

00...

±

2000

M

Mode.P

.Parameter.M

.Monitoring.M

.MeasVal.U.U

.U.UpLim

0...

±

2000

2000

M

Mode.P

.Parameter.M

.Monitoring.M

.MeasVal.I.Ipol.L

.LowLim

00...

±

2000

M

Mode.P

.Parameter.M

.Monitoring.M

.MeasVal.I.Ipol.U

.UpLim

0...

±

2000

2000

M

Mode.P

.Parameter.M

.Monitoring.M

.MeasVal.Up

.Upol.L

.LowLim

00...

±

200.0

M

Mode.P

.Parameter.M

.Monitoring.M

.MeasVal.Up

.Upol.U

.UpLim

0...

±

200.0

200.0

M

Mode.P

.Parameter.M

.Monitoring.M

.MeasVal.A

.Action end, hold, wait, none

none

Parameters for DOS: Selection of the measured quantity for measured value
monitoring. Measured values (pH, U and I) are entered in the list of measured
points only when the monitoring is active.
Selection of the measurement input (MeasInput) applies with measured quanti-
ties pH and U. diff. means differential amplifier, see page 215.
With Ipol, the requests for the polarization current in uA (Ipol) and .PolElectrTest
apply.
With Upol, the request for the polarization voltage in mV (Upol) applies. Entry in
steps of 10 mV. .PolElectrTest also applies. If the test for polarized electrodes
is switched on (ON), it is performed on the change from the inactive basic
mode to an active mode.
On limit value violation, an action can be selected:
stop: Abort of the determination.
hold: Interrupt dosing until the sequence is continued by manual switching (&m

$C).

wait: Interrupt dosing until the limit value is no longer violated, then continue

automatically.

3.2.2.

3.2.2.57.

57. M

Mode.P

.Parameter.D

.DosPara.B

.BeginMeas

pH:

0...

±

20.00, init

init

U, Ipol: 0...

±

2000, init

init

Upol:

0...

±

200.0, init

init

M

Mode.P

.Parameter.D

.DosPara.E

.EndMeas

as with .Begin, but OFF instead of init

M

Mode.P

.Parameter.D

.DosPara.U

.UnitMeas

read only

Parameters for DOS: Setting of ramp start and ramp end. init means ramp start
with default measured value.

3.2.2.

3.2.2.58.

58. M

Mode.P

.Parameter.D

.DosPara.S

.SweepTime

0...300

300...999 999

Sweep time for the ramp in s.

3.2.2.

3.2.2.59.

59. M

Mode.P

.Parameter.D

.DosPara.D

.Dyn

pH: 0.01...0.25

0.25...20.00, OFF

U:

1...15

15...2000, OFF

Ipol:

1...66...2000, OFF