2.4. Configuration, key <CONFIG>

751 GPD Titrino

8

meas.input: 1
time interval 7 d
time counter 0 d

service: OFF

next service YYYY-MM-DD

system test report: OFF

If "on" has been set:

Measuring input (1, 2, diff.)

Time interval for validation (1...9999 d)

Time counter (0...9999 d)

Counts the number of days since the last time the

counter was reset.

Monitoring the service interval (ON, OFF)

Monitoring is carried out after the Titrino has been

switched on. If the monitoring responds the message

"Service is due" appears. The message vanishes with

<CLEAR>.

If "on" has been set:

Date of next service (YYYY-MM-DD)

System test report printout (ON, OFF)

With "on" the report of the system test is printed out

after the Titrino has been switched on, see also page

225.

>peripheral units

send to COM1: IBM

send to COM2: IBM

man.reports to COM: 1

balance: Sartorius

Settings for peripheral units

Selection of printer (Epson, Seiko, Citizen, HP, IBM) at

the Titrino COM1

"Epson", for Epson

"Seiko", e.g. for DPU-414

"Citizen", e.g. for iDP 562 RS

"HP" e.g. for Desk Jet types. Place curves always at the

beginning of a page as you cannot have them over 2

pages.

"IBM" for all printers with IBM character set Table 437

and IBM graphics, as well as for the data transmission

to a computer or a data system.

COM of Titrino for the output of manually triggered

reports (1, 2, 1&2)

Manually triggered reports e.g. with <PRINT> .... .

Exception <PRINT><REPORTS>: These reports are

outputted on the COM as defined in the method.

Selection of balance (Sartorius, Mettler, Mettler AT,

AND, Precisa)

Sartorius:

Models MP8, MC1

Mettler:

Models AM, PM and balances with 011,

012, and 016 interfaces

Mettler AT: Model AT

AND:

Models ER-60, 120, 180, 182, FR-200, 300

and FX-200, 300, 320

Precisa:

Models with RS232C interface